IT systems · Practical guide

Validating Metrohm tiamo: tracing the result back to the initial sample weight

A sample weight is corrected in Metrohm tiamo, the determination recalculated and the result transferred. Does the destination link the correct version to the sample? Test methods, calculations and history together. For an existing installation, distinguish a product discontinuation notice from evidence of the actual end of support.

Table of contents

At a glance

How do you validate Metrohm tiamo?

Validate tiamo for the particular workflow, from sample identity and input values through titration method and calculation to correction, review and transfer. Define requirements, assess supplier evidence and test remaining questions against independently expected results. Record the edition and version. Software validation does not by itself establish instrument suitability, the validity of the analytical method or the suitability of the whole laboratory process.

  • Edition, software version, and device connectivity are specifically stated.
  • Weight, titer, formula, and unit must be appropriate for the determination.
  • Reasoned corrections need a comprehensible relationship between old and new versions.
  • Cancellation of an order position and end of support are distinct statements.

tiamo 3.0: Accurately define edition and current status

Metrohm describes Tiamo as software for controlling titration systems, including method management, calculation, and result handling. The version 3.0 release note distinguishes between full (without client/server support) and multi (with client/server support). Functions of other Metrohm products such as OMNIS or MagIC Net cannot be inferred solely due to the same manufacturer.

The product page for tiamo 3.0 full USB, order number 6.6056.302, designates this item as discontinued but still supported. It provides no support end date. This is relevant for existing installations: specific information regarding your product and contract is required, not a support cessation date inferred from the discontinuation notice.

The article references Release Notes from November 11, 2020, and the tiamo 3.0 assessment from September 2020. The still linked tutorial dates back to January 2009. These documents are classified according to their respective versions; they do not represent the latest patch or a current release of any operating system.

Sources: [1] [5] [2] [3] [4]

Define computerized system, measurement chain, and analytical method

For a GMP-relevant determination, the question "Is tiamo running?" is insufficient. Describe which sample is being tested, what decision the result supports, and what data are required to support it. A development testing workstation may have different requirements than the same software version in quality control.

This guide addresses human medicinal product GMP. Annex 11 is a GMP guideline for computerized systems. The applicability of FDA Part 11 depends on the relevant record-keeping obligations and actual electronic use. Neither product promotion nor a manufacturer's certificate determines the scope of applicability.

Software validation, device suitability, and analytical method validation address distinct questions. Software may correctly perform computational processing of an entered mass, even if the balance is inappropriate or the sample is misidentified. Chapter 3 covers appropriate measuring devices and their control; Chapter 6 addresses testing methods and calculations. Integrate these responsibilities into your evidence-of-verification concept, without inferring the overall process suitability from a passed software validation.

Sources: [6] [7] [11] [12] [10]

Follow the sample from receipt to the stored result.

The Metrohm tutorial distinguishes commands for titration, calculation, reporting, and database storage. This structure highlights a practical control point: a result may be correctly calculated yet incorrectly assigned to the wrong data set or incompletely transferred.

Therefore, provide a brief procedural description of the method used. Identify the source of sample identification, initial weighing, and other input parameters. Document where the result is stored, reviewed, and optionally transferred. The table below is an independent work aid; it does not replace a complete method specification or device instruction.

Own work aid: verify the determination as a continuous data path
StepTechnical RequirementAppropriate comparison with expected results
Probe and input The identifier and sample weight belong to the correct sample Predefined test sample versus stored assignment
Titration The intended method employs appropriate equipment and input parameters. Method specification against actually used configuration
Calculation Formulas, factors, units, and rounding are correct Independent invoice against stated result
Storage and Correction Determination and relevant history remain assignable Compare original and corrected version
Report and Handover The professionally assessed result is clearly communicated. Verify the version, value, and unit at the recipient end.

Sources: [4] [12]

Verify targeted weighing, titer, and equipment calibration

In a calculated determination, several parameters contribute together. The tutorial presents a formula incorporating endpoint volume, titrant concentration, titer, and sample weight. For your method, determine which values are measured, entered, maintained, or manually input. A formula review should not obscure the source of these values.

For an attached balance, sample assignment and unit selection are key validation considerations. Use a known test value and verify the accepted content within the dataset. The existence of a technical connection does not guarantee that the correct single balance will be used during the next sample change. Planned failure scenarios may include missing transfer or an incorrect unit value.

Determine also how valid reagent and sensor data enter the process. The software can only work with values that are actually used. Scientific validity and device control require appropriate documentation. For critical manual entries, Annex 11 mandates an additional accuracy check; the implementation is selected based on the specific consequences of potential errors.

Sources: [4] [12] [11] [7]

Justify correction and retain the new result version together

A documented erroneous weighing may require a controlled correction. The objective is to achieve a technically accurate and traceable result. There is neither a blanket prohibition against any correction nor authorization to modify entries continuously until a desired value appears.

The system assessment report for tiamo 3.0 describes new versions with saved changes to methods or regulations. In the case of recalculations, previous result versions are to be preserved. The tutorial illustrates this process using a changed scale. However, the existence of this function does not yet answer whether your user rights and test runs are properly set up.

For internal validation, provide a dataset containing a known error and a documented justification for correction. Subsequently, compare the original input, corrected value, result version, and associated report. Additionally, verify who is authorized to perform and professionally assess the correction. Modifying the calculation rule constitutes a separate task from correcting a single erroneous input.

Sources: [3] [4] [8]

Further reading:

Test example: Independent setpoint correction for single-pan balance

This is a worked example. It does not describe a test carried out in tiamo. For a 1:1 calculation, R in mg/g is given by V × c × t × M / m. V is in ml, c in mol/l, M in g/mol and m in g. The factor t is dimensionless. The units first give mg of analyte and then mg/g of sample. The example does not prescribe a chemical test method.

The input dataset T-101 uses V = 10.00 ml, c = 0.1000 mol/l, t = 1.000, and M = 100.00 g/mol. The calculated amount thus amounts to 100.00 mg. For an entered mass of 2.0000 g, the result is R = 50.00 mg/g. A documented entry error is corrected to 2.5000 g; the target value for the revised result version is 40.00 mg/g.

For this work aid, the allowable weighing range is predefined as 1.0000 to 3.0000 g inclusive of the limits. Results are rounded to two decimal places. Each table row starts with a fresh copy of the defined input dataset; titration variation constitutes a separate variant case. Error handling is a self-defined requirement and not a claimed tiamo standard setting.

Worked table of expected calculations and error cases
Input or modificationExpected result calculated in advanceAdditional inspection criterion
Starting charge: m = 2.0000 g; all other values as described 100.00 / 2.0000 = 50.00 mg/g Identification T-101 and employed method match
Corrected entry: m = 2.5000 g 100.00 / 2.5000 = 40.00 mg/g Original status and new result version remain traceable
Own variant case: m = 2.5000 g; t = 0.980 98.00 / 2.5000 = 39.20 mg/g Titre source and value used are correct
Lower limit: m = 1.0000 g 100.00 mg/g; Input within the example range The limit value is handled in accordance with the specified instruction.
Upper limit: m = 3.0000 g 33.33 mg/g; Input within the example range Rounding without changing the original values
m = 0 g or m = 0.9999 g No professionally accepted result: zero incidence or breach of defined limits Defined error handling is applied; no undetected substitute value
After correction, 50.00 mg/g is still being transferred instead of 40.00 mg/g Transfer does not meet the requirement Identify version mix-up prior to use
Routine account without correction rights attempts to correct the entry No effective change made through this account The specified permission boundary is respected

Sources: [4] [12] [7]

Audit Trail, System Log, and Technical Review Distinguish

The manufacturer's report separates the audit trail from the standard technical system action protocol. This distinction is helpful for review: a communication error and a change to a specification address different questions. Actively search for evidence supporting the specific action in question.

The assessment also specifies client time as the time basis. Therefore, a reliable time assignment must be established for cross-workstation comparisons. Define how a reviewer links the correction of T-101 to the user, reason, original information, and new result. These are requirements for your documentation pathway, not assertions of identical mandatory fields across each screen.

A recorded change history does not inherently entail a professional assessment of the change. Identify relevant review triggers and responsibilities. When electronic signatures are used, their significance and assignment must be evaluated separately. The presence of a product function does not imply a universal signature requirement for every interaction nor automatically ensures conformity of the entire process.

Sources: [3] [7] [8]

Report, export, and restore at the same dataset to be verified

For T-101, the professionally assessed version should also be identifiable at the intended target. Select a specific report or export pathway and compare sample identification, result, unit, and version reference. The product page lists export options; this does not constitute a fully validated, customer-specific LIMS interface.

A PDF can support the review process. However, it is not inherently the complete original dataset or protected against modifications due to its format. Define which data are relevant for decision-making and where they remain accessible. The visible number in the report may be insufficient without corresponding input values and a complete data trajectory.

Use a corrected determination also for a restoration scenario. Afterward, assess the relationship between the original and new version, method, and relevant history. Define the required backup scope in coordination with IT and laboratory teams; determination data and administrative information may require different backup responsibilities. A successfully completed backup run alone does not confirm technically viable restoration.

Sources: [1] [3] [8] [7]

Further reading:

Evaluate operational continuity and supplier documentation substantively.

An discontinued order position does not automatically render an existing installation unsuitable. Verify support status, operating system release, drivers, technical protection measures, and availability of required data. The operating system list from the 2020 Release Notes serves as a historical version record and does not constitute current security or support commitments.

For continued operation, the team requires a clear decision regarding its actual risks. Missing manufacturer releases or technically unsustainable dependencies may trigger action. A future migration will be planned with regard to data volume, readability, and intended use; it should not simply defer unresolved data retention to a new system.

Evaluate existing supplier and service provider documentation based on the scope of the most recent review. Specific tests do not need to be routinely repeated. A general assessment does not replace the need to verify your calculation rules or role definitions. Maintain a consolidated record of transfer decisions, additional cases, and remaining restrictions such that future use can be professionally assessed.

Sources: [1] [2] [3] [7]

What Laboratory, Quality Assurance, and IT should provide prior to commencement

The laboratory supplies the methods used, expected values and suitable sample data. QA clarifies the requirements and how corrections are handled. IT describes installation, access, backup and recovery. Involve Metrohm or the support partner where product capability or support coverage remains unclear. Give each open question an owner.

The effort depends particularly on different calculations, device integrations, existing documentation, and data transfers. Even a single workstation can be demanding if its method or history is unclear. A one-size-fits-all project duration based solely on product name or company size would obscure these differences.

  • Compile edition, build, device, drivers, and current support information.
  • Name the methods, formulas, units, and rounding rules used.
  • Clear sample assignment as well as origin of weighing, titer and other input sizes.
  • Define rights and procedural requirements for correction and re-evaluation.
  • Prepare test data sets with independent expected results, failure cases, and defined initial states.
  • Assign reports, exports, backup scope, and recovery documentation.

Sources: [9] [7]

Project experience and collaboration

Link the titration calculation to the original result

With laboratory owners, we trace a titration from sample weight and titre to the stored and reviewed result. A known calculation also lets us examine a justified correction and the link between the original and revised result. A narrowly defined tiamo use may need a focused test plan. Changes to the measurement chain, analytical method or data handover need evidence agreed with the relevant owners. A software test cannot resolve an outstanding assessment of the analytical procedure.

Frequent Questions on tiamo and Titration Results

Is Tiamo 3.0 retired or still supported?

The manufacturer's page for tiamo 3.0 full USB, order number 6.6056.302, states that the product has been discontinued and is no longer supported. No support contact is available on this page. Therefore, please clarify the current scope of support and contract terms for your installation and agreement. The notice shall neither be interpreted as an unlimited support guarantee nor as a mandatory obligation to immediately cease operation.

Sources: [1]

What distinguishes tiamo full from multi?

The version 3.0 vendor statement specifies "full without client/server support" and "multi with client/server support." This limitation impacts architecture, administration, and data retention. Verify the actually licensed edition and installation. No general conclusions can be drawn regarding light, other versions, or the full range of features of further Metrohm products.

Sources: [5]

Does the supplier assessment constitute validation?

It serves as a source for the described functions and responsibilities. In assessing your documentation, you shall additionally evaluate its alignment with version, configuration, method, and usage. Suitable evidence may be adopted; open requirements require supplementation. The report itself assigns system validation responsibility to the operator. A manufacturer's self-certification is not an independent FDA certificate for your laboratory.

Sources: [3] [7]

Does tiamo validation replace analytical method validation?

No. The system validation may include inputs, calculations, and data storage. The analytical method must be appropriate for its intended use; likewise, measuring instruments require appropriate controls. An incorrect software version may arise from an inappropriate measurement. Therefore, assign the evidence to their respective questions and link them for the overall technical evaluation.

Sources: [12] [11]

May an incorrect sample weight be corrected later?

A justified correction may be required. It must follow a defined procedure with a clear rationale and appropriate authorizations. The tiamo-3.0 report describes the retention of previous versions during recalculation. Review this function and the technical assessment within your workflow. The correction must not constitute an ungrounded alteration made to achieve a desired outcome.

Sources: [3] [8]

How do we verify our calculations independently?

Prepare input values and setpoints outside the scope of the to-be-verified calculation. Consider units, titers, rounding, and professional limits. Also include values where a swapped factor or incorrect unit is evident. The result must not be finalized as the desired value only after reviewing the program output. The number of cases follows relevant risk assessments.

Sources: [12] [7]

Are Audit Trail and System Log the same?

The manufacturer's report for tiamo 3.0 explicitly distinguishes between these. Technical events may support, for example, an investigation into a communication failure. A change to a specification requires the corresponding change justification. Define which information your technical assessment requires and how it will be consolidated. A large volume of technical alerts does not automatically validate the required data history.

Sources: [3]

What must be available prior to project start and who is involved?

The following are required: installation overview, methods and calculations, existing records, and a described sample-to-result pathway. The laboratory determines the intended reference values, QA defines the evidence requirements, and IT handles the technical operational responsibilities. Manufacturers or partners provide product boundaries. For a small team, clearly defined responsibilities are more effective than multiple disjointed document lists.

Sources: [7] [12]

What factors drive effort and duration?

Multiple method variants, manual entries, custom exports and unresolved legacy data increase the work. Relevant existing evidence and prepared test data help. Allow time for laboratory decisions and access to a suitable test environment. The designation “tiamo 3.0” alone does not establish a reliable price or duration.

Sources: [9] [7]

Can we only transfer PDFs during a system change?

For dynamic analytical data, a static PDF report alone is insufficient. When input parameters and calculations can be reapplied, the required relationship between original data, processing, and results must be preserved. The FDA outlines this limitation for CGMP laboratory records in its Data Integrity Guidance. A PDF can supplement the presentation of results. Define the required retention period and assess access controls as well as future readability following relevant technical changes.

Sources: [8] [7] [3] [13]

Sources and review date

Status as of October 6, 2026. Sources include tiamo 3.0, Release Notes 11.11.2020, and Assessment 09/2020; the linked tutorial originates from 01/2009. Current platform access permissions and support scope require specific clarification. The computational example is an internal reference tool and does not correspond to a real sample or completed product test. Additional targeted source cross-reference on October 18, 2026: FDA Data Integrity Guidance 2018, Questions 1d and 10, relating to dynamic laboratory records.

  1. tiamo 3.0 full USB: 1 license — 6.6056.302
  2. Release Notes tiamo 3.0
  3. tiamo 3.0 — System Assessment Report
  4. tiamo Tutorial — linked on the product page as tutorial tiamo 2.X and 3.X
  5. tiamo 3.0 full / multi — Declaration of Conformity
  6. European Commission: EudraLex Volume 4
  7. EU GMP Annex 11: Computerised Systems
  8. EU GMP Chapter 4: Documentation
  9. EMA: ICH Q9(R1), Step 5, Corr.2
  10. FDA: Part 11 — Scope and Application
  11. EU-GMP Chapter 3: Facilities and Equipment
  12. EU-GMP Chapter 6: Quality Control
  13. FDA: Data Integrity and Compliance With Drug CGMP – Questions and Answers, December 2018Questions 1d and 10 regarding dynamic laboratory records on 18.10.2026 read in original. Scope: Medicinal Products – CGMP; PDF may serve as supplementary results presentation.

Sources checked: .

Daniel Herrmann, Daniel Herrmann Consulting

Daniel Herrmann Consulting